Clinical Accuracy Validation of the Moni-Patc Temperature Monitoring Patch During Surgery (Compared With Esophageal and Bladder Temperature)
For patients and families
In plain language
An automatic summary of structured registry data. It is an orientation aid, not a substitute for the official protocol or a physician assessment.
- What is being studied
- The protocol lists: Moni-Patc Temperature Monitoring Patch System.
- Who it may be relevant to
- Registry conditions: Temperature Monitoring, Intraoperative. Basic parameters: from 18 years · All.
- What needs checking
- Age, condition and sex are only basic indicators. Prior treatment, laboratory values and other mandatory requirements appear in the eligibility criteria below.
- Where it takes place
- Center list to be confirmed — check the primary protocol.
- Next step
- Save the trial, show it to the treating physician, and confirm current recruitment with the study center. Costs, documents and travel →
Unsure about the terms? Read our patient guide →
Official title
Moni-Patc Temperature Monitoring Patch System Clinical Accuracy Validation
Overview
Moni-Patc Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck.This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous esophageal and bladder temperature monitoring device used as the clinical reference.Main ObjectivesTo verify that the Moni-Patch and the reference device(esophageal and bladder) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C).To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.
Detailed description
This single-center, open-label accuracy validation compares continuous core temperature measured by the Moni-Patch applied on the neck with esophageal (T\_eso) and bladder (T\_bld) temperatures during elective surgeries under general anesthesia. Approximately 30 adult participants will be monitored for ≥60 minutes intraoperatively. Agreement will be assessed using Bland-Altman methods (bias and 95% limits of agreement). The device is non-invasive and Bluetooth-enabled; readings are captured via the receiver and routed to the OR multi-parameter monitor and EMR. Safety monitoring focuses on device-related skin reactions and device deficiencies; no changes to anesthetic care are mandated by the protocol.
Interventions
- Device Moni-Patc Temperature Monitoring Patch System
A non-sterile, non-invasive temperature monitoring system applied to the neck; Bluetooth 5.0 transmission to a receiver; continuous core temperature estimation using dual heat-flux/thermistor-based algorithm; conforms to ISO 80601-2-56 (25-42 °C).
Primary outcome measures
- Agreement Between Moni-Patch™ Temperature Monitoring Patch System and Esophageal (Teso) and Bladder (Tbld) Core Temperature Measurements [Time frame: Continuous intra-operative monitoring for ≥60 minutes, beginning after ≥30 minutes of Moni-Patch sensor acclimation prior to operating room entry.]
Eligibility criteria
Inclusion criteria
- Age ≥18 years
- Subject is undergoing an elective surgical procedure expected to require at least 60min of general anesthesia and during which routine esophageal (Teso) and bladder (Tbld) temperature monitoring is expected.
- Subjects are expected to remain hospitalized at least overnight after surgery.
- Subject specifically agrees to temperature Sensor placement on the neck prior to surgery.
- Must be able to attach the Sensor away from the operative field (neck) with uninterrupted wireless connection between the Sensor and the Receiver (as close as possible).
- Use environment must meet both of the following parameters:
- Temperature 15-40⁰ C/59-104⁰ F
- Humidity 30-75% RH (non-condensing)
- For non-English speaking subjects, a translator must be present at the time of consent.
Exclusion criteria
- Subjects unable to provide their own consent. Consent from family or LOR is not acceptable.
- Subjects undergoing surgical procedures who have conditions of the esophagus or bladder that contraindicate placement of these monitoring probes/sensor.
- Subjects whose surgery does not involve the monitoring of esophageal or bladder temperature.
- Subjects undergoing surgery that prohibits placement of an esophageal temperature monitoring probe/sensor (e.g. neck surgery).
- Subjects with medical conditions such as inflammation, scratches, redness, rash at the Sensor placement site.
- Subjects for whom the sensor cannot be securely adhered to the skin without floating or detaching, or accurately positioned over the carotid artery.
- Not for use in MRI or CT facilities.
- Use environment does not meet one or more parameters.
Criteria are shown verbatim from the registry (in English). Final eligibility is always assessed by the study center.
Healthy volunteers: No
Study design
- Observational model
- Case-only
Study locations
Center list to be confirmed — check the primary protocol.
Identifiers
NCT: NCT07397065 · MM-MPS2-0001