Menu
Recruiting NCT06552689

Sutra Hemi-valve First-in-Human Study

No phase Interventional Mitral Regurgitation

For patients and families

In plain language

An automatic summary of structured registry data. It is an orientation aid, not a substitute for the official protocol or a physician assessment.

What is being studied
The protocol lists: Sutra Hemi-valve Transcatheter Mitral Valve Replacement (TMVR) System.
Who it may be relevant to
Registry conditions: Mitral Regurgitation. Basic parameters: from 18 years · All.
What needs checking
Age, condition and sex are only basic indicators. Prior treatment, laboratory values and other mandatory requirements appear in the eligibility criteria below.
Where it takes place
New Zealand
Next step
Save the trial, show it to the treating physician, and confirm current recruitment with the study center. Costs, documents and travel →
Official title

First Human Study of Sutra Hemi-Valve to Evaluate Safety and Efficacy in Patients With Mitral Regurgitation

Overview

Study to evaluate the feasibility, safety and performance of the Sutra Hemi-valve Transcatheter Mitral Valve Replacement System in patients with moderate-severe and severe symptomatic mitral regurgitation who are at high risk for surgical treatment.

Detailed description

Investigational Device:

Sutra Transcatheter Mitral Valve Replacement (TMVR) System The Sutra Hemi-valve TMVR System consists of 32Fr delivery devices and an implantable hemi-valve.

The Sutra Hemi-valve consists of a self-expanding nitinol frame, a tri-leaflet bovine pericardial hemi-valve, and a sealing skirt. The Sutra Hemi-valve is intended to replace the posterior leaflet of the mitral valve to treat mitral regurgitation. The frame consists of an atrial portion for anchoring to the native annulus, and a ventricular portion which houses the Hemi-valve. The atrial section of the Sutra Hemi-valve contains three through-holes for anchoring to Dual-guiding-and-fixation (DGF) devices implanted at the lateral trigone (T1), the medial trigone (T3), and in the center of the posterior annulus (P2). The Sutra Hemi-valve comes in two sizes: Size 42 and Size 38.

Intended Use: The Sutra Hemi-valve TMVR system is intended for use as a partial replacement mitral valve.

Indication for Use: The Sutra Hemi-valve TMVR system is indicated for use in patients with moderate-severe or severe symptomatic mitral regurgitation who are considered high risk for conventional mitral valve surgery.

Trial Design: Prospective, multi-center, unblinded, single arm first in human/ feasibility clinical study.

Primary Objective:

To assess the safety and performance of the Sutra Hemi-valve TMVR System in patients with moderate-severe or severe symptomatic mitral regurgitation who are considered high risk for conventional mitral valve surgery.

Secondary Objectives:

* Improvement of mitral regurgitation * Assess interaction of implant prosthesis with native mitral valve apparatus, conduction system, left ventricular outflow tract (LVOT) and aortic valve * Assess stability of implant. * Assess hemodynamics of the mitral valve apparatus

Interventions

  • Device Sutra Hemi-valve Transcatheter Mitral Valve Replacement (TMVR) System
    Trans-septal Sutra Hemi-valve TMVR System to treat mitral regurgitation

Primary outcome measures

  • Primary Safety Endpoint - The ability of the Sutra Hemi-valve to be placed without Major Device Related Adverse Events through thirty (30) days [Time frame: Day 30]
  • Primary Performance Endpoint - Sutra Hemi-valve TMVR System Technical Success [Time frame: Day 0]
Secondary outcome measures (5)
  • Intra-procedure MR severity is reduced by at least 1 grade from baseline per echo (e.g., from severe to moderate). [Time frame: Day 0]
  • Post-index procedure (discharge or within 7 Days post-index procedure, whichever is earliest) MR severity is reduced by at least 1 grade from baseline per echo (e.g., from severe to moderate). [Time frame: Day 0 - Day 7]
  • MR severity grade per Echo [Time frame: Day 30, Month 6, Month 12]
  • Change in NYHA functional class [Time frame: Day 30, Month 6, Month 12]
  • 6 Minute Walk Test distance (6MWT) [Time frame: Day 30, Month 6, Month 12]

Eligibility criteria

Inclusion criteria

  • Subject is greater than 18 years of age at time of enrollment.
  • Subject has moderate-severe or severe symptomatic mitral regurgitation who are considered high surgical risk for conventional mitral repair or replacement, as assessed by institutional multidisciplinary heart team.
  • New York Heart Association (NYHA) Functional Class II and above.
  • Subject anatomy is suitable for transfemoral and transeptal device delivery and meets sizing criteria for Sutra Hemi-valve TMVR System.
  • Subject willing and able to provide informed consent and able to comply with the study protocol and follow up.
  • Female subjects of childbearing potential have a negative pregnancy test ≤ 7 days before the procedure

Exclusion criteria

  • Prohibitive mitral annular calcification
  • Diseased mitral anterior leaflet such as flail or prolapse
  • Previous mitral valve intervention
  • Implanted with venous stents (iliac and femoral) including inferior vena cava (IVC) filter or congenital abnormalities of the IVC that would preclude ability for transfemoral access with the delivery system.
  • Evidence of intra-cardiac, inferior vena cava (IVC) or ipsilateral femoral venous thrombus
  • Severe aortic stenosis or insufficiency (Note: may be treated ≥ 30 days prior to study procedure)
  • Contraindication for transesophageal echocardiography (TEE) or MDCT scan
  • Active infections requiring current antibiotic therapy (if temporary illness, patients may enroll 2 weeks after discontinuation of antibiotics)
  • Endocarditis within 6 months
  • Left ventricular ejection fraction (LVEF) < 25%
  • Severe Chronic Obstructive Pulmonary Disease (COPD) demonstrated by Forced Expiratory Volume (FEV1) < 750cc
  • Implant or revision of any pacing device < 30 days prior to intervention
  • Symptomatic coronary artery disease treated < 30 days prior to study procedure
  • Active peptic ulcer or upper gastrointestinal bleeding within 90 days
  • Prior stroke, TIA, or myocardial infarction within 90 days
  • Severe renal insufficiency (creatinine > 225 µmol/L) or patient requiring dialysis
  • Any endovascular therapeutic interventional or surgical procedure performed within 30 days prior to enrollment
  • Pregnant, nursing or planning to be pregnant. (Female participants of childbearing potential must have a negative pregnancy test prior to enrollment)
  • History of bleeding diathesis or coagulopathy that cannot be managed or patient will refuse blood transfusion
  • Hemodynamic instability requiring dependency on either inotropic agents or mechanical circulatory support
  • Known hypersensitivity or contraindication to heparin and bivalirudin
  • Known allergy to nitinol or contrast agents that cannot be pre-medicated
  • Severe dementia or lack of capacity due to conditions that result in either inability to provide informed consent for the trial/procedure, prevent independent lifestyle outside of a chronic care facility, or will fundamentally complicate rehabilitation from the procedure or compliance with follow-up assessments.
  • Patient is participating in another investigational drug or device clinical study that has not completed primary endpoint follow-up

Criteria are shown verbatim from the registry (in English). Final eligibility is always assessed by the study center.

Healthy volunteers: No

Study design

Allocation
N/A
Model
Single group
Masking
Open label
Primary purpose
Treatment

Study locations

New Zealand · 1 center
  • Waikato Hospital — Hamilton

Identifiers

NCT: NCT06552689 · CIP-002

Primary sources (government registries)