Metal-ions in Patients With the PEEK HD Coupling Mechanism of the Knee
For patients and families
In plain language
An automatic summary of structured registry data. It is an orientation aid, not a substitute for the official protocol or a physician assessment.
- What is being studied
- The protocol lists: Blood sample analysis.
- Who it may be relevant to
- Registry conditions: Mutars, Modular Universal Tumor And Revision System, Metal-ions, Tumor Prosthesis. Basic parameters: from 18 years · All.
- What needs checking
- Age, condition and sex are only basic indicators. Prior treatment, laboratory values and other mandatory requirements appear in the eligibility criteria below.
- Where it takes place
- Netherlands
- Next step
- Save the trial, show it to the treating physician, and confirm current recruitment with the study center. Costs, documents and travel →
Unsure about the terms? Read our patient guide →
Official title
Systemic Metal Ion Concentrations in Patients With a PEEK HD Coupling Mechanism in Modular Mega-prostheses of the Knee; a Prospective Cohort Pilot Study
Overview
The goal of this observational study is to evaluate the changes of serum metal ion (Cobalt and Chrome) levels at 0, 3-6, 12, and 24 months postoperative in patients receiving a MUTARS mega-prostheses of the knee with a PEEK HD coupling mechanism. The main question(s) to answer are: • \[what is the change in serum level metal ions (Cobalt and Chrome) in patients receiving a primary MUTARS knee endoprosthesis with the PEEK HD coupling mechanism \] • \[what is the change in serum level metal ions (Cobalt and Chrome) in patients with a MoM coupling mechanism revised to the PEEK HD coupling mechanism\]. Participants will be asked to fill out functional outcome measures during the period under study.
Detailed description
Rationale:
Local release of metal ions and their systemic sequelae is an increasing source of concern in patients with an endoprosthetic reconstruction of major joints. It is assumed that metal-on-metal (MoM) articulations, used in several types of mega-prostheses, may cause release of metal ions. Corrosion of non-articulating surfaces, abrasive wear of soft tissues and fretting of modular junctions could contribute to this release. The release of metal ions are known to induce inflammatory responses and immune reactions in the directly exposed tissues and can cause serious local adverse reactions such as metallosis, osteolysis, pseudotumor formation and systemic adverse effects such as cardiovascular and neurological adverse effects. Our study group previously found significantly elevated serum levels of Silver (Ag), Chrome (Cr) and Cobalt (Co) in a series (not published yet) of 11 patients with a MoM coupling mechanism of knee endoprostheses. Furthermore, a number of early mechanical failures of these MoM coupling mechanisms were encountered by the investigators recently.
The implant manufacturer (ImplantCast GmbH) recently introduced a carbon-reinforced PEEK HD (polyether ether ketone high demand) coupling mechanism in an attempt to reduce the risk of early mechanical failure and to lower the risk on the release of metal ions. Based on the outcomes of previous research and the mechanical failures of the MoM coupling mechanism, the investigating center will switch to the use of the new (approved and CE-marked) PEEK HD coupling mechanism. Hypothesized is that the PEEK HD coupling mechanism will not result in elevated serum metal ion levels in patients receiving a mega prosthesis.
Objective:
The aim is to 1) monitor and investigate the change in serum metal ion (Cobalt and Chrome) levels postoperatively (at 0, 3-6, 12 and 24 months) in patients receiving a primary PEEK HD coupling mechanism, monitor and investigate the change in serum metal ion levels pre-and postoperatively in patients with an MoM coupling mechanism which is revised to a PEEK HD coupling mechanism (in case of failure for any reason), 2) document adverse effects caused by metal ions, 3) report factors (possibly) correlating with metal ion release and 4) evaluate the functional outcome of the patient using the Patient-Reported Outcome Measurement Information System (PROMIS) 29 and or Toronto Extremity Salvage Score (TESS) lower extremity.
Study design:
This is a prospective cohort pilot study to evaluate metal ion concentrations, possible adverse effects and functional outcomes after the implantation of a PEEK HD coupling mechanism.
Study population:
Patients ≥18 years receiving a reconstruction with a MUTARS knee endoprosthesis (distal femur, proximal tibia or total knee) with a PEEK HD couplingmechanism, and patients receiving a PEEK HD coupling mechanism in the case of a revision procedure of a MUTARS knee endoprosthesis with a MoM coupling mechanism.
Main study parameters/endpoints:
(1) Serum metal ion concentrations Cobalt (Co) and Chrome (Cr) pre- and postoperatively (at 3-6, 12 and 24 months), (2) Secondary parameters are used to identify any adverse effects in patients with increased serum levels of ion metals (metallosis, osteolysis, periprosthetic loosening, pseudotumor formation). (3) Tertiary parameters are used to identify possible risk factors correlating with metal-ion release. (4) quaternary parameters (PROMIS 29 or TESS) are used to evaluate the functional outcome of the patient. All other variables (e.g. general baseline characteristics, treatment characteristics, and surgical details including prosthesis details) will be gathered within the "MORE" cohort study and "biobank" study (study number 2022-031, reference BWD005/SH/sh).
Nature and extent of the burden and risks associated with participation, benefit and group relatedness:
Serum metal ion concentrations (Cobalt and Chrome) will be determined at 0, 3-6, 12, and 24 months postoperatively. The potential benefits of routine determination of metal ion concentrations are early detection of toxic values of metal ions. Except for the negligible risks of routine venepuncture no potential risks are anticipated.
Interventions
- Other Blood sample analysis
Blood sample analysis on Cobalt and Chrome will be performed during the period under study.
Primary outcome measures
- Change in serum metal ion levels (Cobalt and Chrome) [Time frame: 24 months]
Secondary outcome measures (1)
- Adverse effects [Time frame: 24 months]
Eligibility criteria
Inclusion criteria
In order to be eligible to participate in this study, a subject must meet all of the following criteria: The patient
- is 18 years of age or older
- receives a MUTARS knee replacement with PEEK HD coupling mechanism, or undergoes a revision (for any reason) of a MUTARS knee replacement during which the MoM coupling mechanism is revised for a PEEK HD coupling mechanism
- Is able to give informed consent
Exclusion criteria
A potential subject who meets any of the following criteria will be excluded from participation in this study:
- Anamnestic use of metal containing nutritional supplements or medications
- Contact with metal ions in the work environment
- Renal insufficiency defined as an eGFR<60
- Presence of implants containing Cobalt and Chrome (including non-orthopaedic implants such as stents and dental implants)
Criteria are shown verbatim from the registry (in English). Final eligibility is always assessed by the study center.
Study design
- Observational model
- Cohort
Study locations
Netherlands · 1 center
- Leiden University Medical Center — Leiden
Identifiers
NCT: NCT05893602 · NL82185.058.22